Validation of Konsung Compass 2000 Dry Biochemical Analyzer

  • Yingchun Li Jiangsu Konsung Bio-Medical Science and Technology Co., Ltd.
  • Xue Zhang Jiangsu Konsung Bio-Medical Science and Technology Co., Ltd.
  • Tong Chen Jiangsu Konsung Bio-Medical Science and Technology Co., Ltd.
  • Xiangwei Fan Jiangsu Konsung Bio-Medical Science and Technology Co., Ltd.
  • Haibo Qian Jiangsu Konsung Bio-Medical Science and Technology Co., Ltd.
Keywords: POCT Dry Biochemical Analyzer, Blood Lipid, Blood Glucose

Abstract

Dry biochemical analyzers have been increasingly popular in many tests by primary hospitals, field hospitals and other areas subject to economic and medical underdevelopment as well as poor transportation. With the increasing demand for POCT in primary medical care around the world, upgrading of dry biochemical analyzers has been a hot topic in technical research. Against such context, Konsung Compass2000 dry biochemical analyzer, a POCT system with high precision and accuracy, is developed. Furthermore, the upgraded dry biochemical analyzers can, in a more convenient and accurate way, monitor glucose, lipid and other indices affecting the course of chronic diseases.

References

[1] Guidelines for Risk Assessment and Management of Cardiovascular Disease in China [J]. Chinese Circulation Journal, 2019, 34 (01): 4-28.

[2] Zhao, C, Ren, R, A Probe into Prevention and Treatment Measures for Chronic Diseases [A]. Dalian Medical University, Medicine & Philosophy. Collections of Papers of the Summit Forum on Medicine Sciences & Humanities and the First Meeting of the Fourth Editorial Board of Medicine & Philosophy [C]. Dalian Medical University, Medicine & Philosophy Periodical Office:Medicine & Philosophy of Chinese Society for Dialectics of Nature, 2016: 7.

[3] Dean, D, et al. Development and Evaluation of a Point-of-Care Test (POCT) in a Low Resource Setting with High Rates of Chlamydia Trachomatis Urogenetic Infections in Fiji. [J]. Journal of Clinical Microbiology, 2021.

[4] Lu, F, et al. A Fluorescence Color Card for Point-of-Care Testing (POCT) and its Application in Simultaneous Detection. [J]. The Analyst, 2021.

[5] Duan, XY, Wang, S, Wu, LC. Application and Management of POCT in Emergency Medicine [J]. Laboratory Medicine and Clinic, 2020, 17 (17): 2582-2585.

[6] Guo, ZL. Application of BNP Level Measured by Bedside POCT in Differential Diagnosis of Pulmonary and Cardiogenic Dyspnea [J]. The Journal of Medical Theory and Practice, 2020, 33 (11): 1846-1847.

[7] Bao, J, Yang, YJ, Chen, YM, Liu, H, Cao, LX, Discussion on Current Application Status and Development Prospect of POCT Technology [J]. Medical Innovation of China, 2016, 13 (30): 138-141.

[8] Cheng, SY, et al. Mobile Health for Diagnosis-PixoTest HbA1c POCT System [J]. Journal of Medical and Biological Engineering, 2020, 40 (1): 138-142.

[9] Liu, J, Yao, SP, Expert Consensus on the Application of Point-of-Care Testing (POCT) in Community Medical Institutions [J]. China Medical Equipment, 2019, 16 (08): 143-145.

[10] Van, HV, et al. Organization and Quality Monitoring for Point-of-Care Testing (POCT) in Belgium: Proposal for an Expansion of Legal Framework for POCT into Primary Health Care. [J]. Acta clinicaBelgica, 2021: 1-8.

[11] Shao, MX, Li, SJ, Li, XJ, Cao, WJ, Performance Evaluation of Huizhong MQ-8000 Glycosylated Hemoglobin Analyzer [J]. Labeled Immunoassays and Clinical Medicine, 2021, 28 (07): 1249-1252.

[12] Jing, H, Sun, J, Li, D, Jing, B, Ma, YX, Wang, SH, Performance Evaluation of Four POCT Glucose Meters and Comparative Analysis on Glucose Test Results of Automatic Biochemical Analyzer [J]. Smart Healthcare, 2021, 7 (14): 8-11 +15.

[13] Ji, YY, Feng, JW, Bai, Y, Huang, AJ, Shen, H, Performance Evaluation of BC-6800plus Automatic Hematology Analyzer [J]. China Medical Equipment, 2021, 18 (02): 20-24.

[14] Gong, Y, Lu, RQ, Guo, L, Performance Evaluation of Roche Modular Automatic Biochemical Analyzer for Determination of Blood Urea Nitrogen, Creatinine and Uric Acid [J]. International Journal of Laboratory Medicine, 2011, 32 (11): 1225-1226 +1230.

[15] Feng, Q, Deng, DY, Zhou, LH, Chen, D, Li, ZA, Zhi, GH, Xue, YS. Evaluation of Analytical Performance of Automatic Biochemical Analyzer for Determination of Serum LDL Cholesterol [J]. Jilin Medical Journal, 2013, 34 (13): 2407-2409.

[16] Zhao, MY, Wang, L, [J]. Comparative analysis of the same test items between blood gas analyzers and fully automatic blood analyzers/fully automatic dry biochemical analyzers [J]. Biomedical Engineering and Clinical Medicine ,2022, 26(01):104-107.

[17] Gao, Y, Xiao, LJ, Analytical validation of the detection performance of Johnson & Johnson V-350 Fully Automatic Dry Biochemical Analyzer [J]. China Medical Device Information, 2021, 27(15): 22-24.

[18] Chen, JH, Yang, GL, Huang, Y, Performance validation of a fully automatic dry biochemical analyzer [J] Laboratory Medicine and Clinic. 2012,9(23).
Published
2022-07-07
Section
Original Research Article